Show categories
Measurement method | Tuning fork vibro method (natural frequency at 30Hz) | |||
Amplitude range | 0.07 to 1.2 mm | |||
Viscosity measurement range | amplitude | range | ||
0,07 mm | 2,000 to 25,000 mPa·s | |||
0.1 mm | 20 to 25,000 mPa·s | |||
0.2 mm | 0.3 to 25,000 mPa·s | |||
0.4 mm | 0.3 to 12,000 mPa·s | |||
0.6 mm | 0.3 to 5,000 mPa·s | |||
0.8 mm | 0.3 to 5,000 mPa·s | |||
1.0 mm | 0.3 to 3,000 mPa·s | |||
1.2 mm | 0.3 to 3,000 mPa·s | |||
Repeatability | 1% of the measured value (standard deviation) | |||
Accuracy | ±3% (1 to 1000 mPa.s) when the amplitude is set to 0.4 mm | |||
Minimum display | Range (mPa·s) | Minimum display (mPa·s) | Minimum display (Pa·s) | |
0.3 to 10 | 0.01 | 0.0001 | ||
10 to 100 | 0.1 | 0.0001 | ||
100 to 1000 | 1 | 0.001 | ||
1000 to 25000 | 10 | 0.01 | ||
Viscosity measurement unit | mPa·s, Pa·s, cP, P | |||
Minimum sample amount | 10 ml | |||
Temperature display | 0 to 99°C/0.1°C ; 100 to 160°C/1°C |