UV-VIS Spectrophotometer Shimadzu UV-1280

UV-VIS Spectrophotometer Shimadzu UV-1280

Order Number: 0205.UV1280
Designed by the leaders in UV-Visible Spectroscopy for molecular absorption quantitative analysis, the UV-1280 Multipurpose UV-Visible Spectrophotometer offers wavelength scanning from 190-1100 nm.
This lower-cost, high-quality instrument is ideal for applications ranging from routine environmental and food quality testing to life science analyses.
More about the product

Product is on order, write to us at info@helago-cz.cz or fill in the form.

More than a purchase. You get service and expert advice. Ask which products and combinations are recommended for you.

Spectral bandwidth: 5 nm
Wavelength range: 190.0 to 1100.0 nm
Wavelength display: 0.1 nm increments
Wavelength setting: 0.1 nm increments (1 nm increments when setting scanning range)
Wavelength accuracy: ±1.0 nm
Wavelength repeatability: ±0.3 nm
Wavelength slew rate: Approx. 6,000 nm/min, Wavelength scanning speed: Approx. 9 nm/min to 1,600 nm/min
Light source switching:

Select from the following three:

  • Automatic switching linked to the wavelength. The switching wavelength can be selected between 295 nm and 364 nm, in 1 nm increments. The recommended wavelength is 340 nm.
  • Use the halogen (WI) lamp only, with no switching
  • Use the deuterium (D2) lamp only, with no switching

Stray light: 0.05 % or less (220.0 nm NaI, 340.0 nm NaNO2)
Photometric system: Monitor double beam optics

Photometric range:

  • Absorbance: -0.3 to 3.0 Abs
  • Transmittance: 0.0 % to 200 %

Recording range:

  • Absorbance: -4.0 to 4.0 Abs
  • Transmittance: -400 % to 400 %

Photometric accuracy:

  • ±0.005 Abs at 1.0 Abs
  • ±0.003 Abs at 0.5 Abs
  • Using NIST 930D filter
UV-VIS Spektrofotometr - Shimadzu UV-1280

UV-VIS Spectrophotometer Shimadzu UV-1280